e-CFR banner

Home
gpo.gov
govinfo.gov

e-CFR Navigation Aids

Browse

Simple Search

Advanced Search

 — Boolean

 — Proximity

 

Search History

Search Tips

Corrections

Latest Updates

User Info

FAQs

Agency List

Incorporation By Reference

eCFR logo

Related Resources

 

Electronic Code of Federal Regulations

e-CFR data is current as of November 7, 2019

Title 47Chapter ISubchapter DPart 95Subpart H → §95.2379


Title 47: Telecommunication
PART 95—PERSONAL RADIO SERVICES
Subpart H—Wireless Medical Telemetry Service


§95.2379   WMTS unwanted emissions limits.

Each WMTS transmitter type must be designed to comply with the requirements in this paragraph.

(a) Unwanted emissions on frequencies below 960 MHz must not exceed 200 μV/m, measured at a distance of 3 meters using measuring instrumentation with a CISPR quasi-peak detector.

(b) Unwanted emissions on frequencies above 960 MHz must not exceed 500 μV/m, measured at a distance of 3 meters using measuring equipment with an averaging detector and a 1 MHz measurement bandwidth.

Need assistance?