PART 95—PERSONAL RADIO SERVICES
Subpart H—Wireless Medical Telemetry Service
Each WMTS transmitter type must be designed to comply with the requirements in this paragraph.
(a) Unwanted emissions on frequencies below 960 MHz must not exceed 200 μV/m, measured at a distance of 3 meters using measuring instrumentation with a CISPR quasi-peak detector.
(b) Unwanted emissions on frequencies above 960 MHz must not exceed 500 μV/m, measured at a distance of 3 meters using measuring equipment with an averaging detector and a 1 MHz measurement bandwidth.